|
|
| |
|
| ::
REAGENT STRIPS FOR URINE ANALYSIS :: |
|
|
|
|
|
 |
| Item
Code: US-01 |
|
|
| HICKS
URISCREEN Reagent Strips (11 Parameters) |
| |
INTENDED
USE
Urianalysis Reagent Strips test for Ascorbic acid, Glucose,
Bilirubin, Ketone, Specific Gravity, Blood, pH, Protein, Nitrite,
Leukocytes. The strip may be lead visually or instrumentally,
using the appropriate Urine Chemistry Analyzers such as the
Clintek family of readers. |
| |
| SPECIMEN
COLLECTION & PROCEDURE |
| 1. |
Use
a fresh urine specimen, less than 4 hours
old, and place it into a clean, dry container.
Do not centrifuge. Test within one hour.
If not possible, refrigerate and restore
to room temperature before testing.
|
| |
|
| 2. |
Remove
one strip from the bottle and replace
the cap lightly. Briefly (no longer than
one second) immerse alt reagent areas
into the specimen. Wipe off excess urine
on the rim of the container.
|
| |
|
| 3. |
Hold
strip in vertical position. Refer to
the bottle label for specific reagent
areas on the test strip. Compare the
test areas with the color scale on the
label. Proper reading times are critical
for optimal results. See each reagent
time as indicated on bottle label. Coloration
appearing only along the edges of the
test, or developing after more than two
minutes, has no diagnostic value.
|
|
| |
LIMITATION
As with all laboratory tests, definitive diagnostic or therapeutic
decisions should not be based on any single test result
or method.
STORAGE
Do not remove desiccants from the bottle. Store at room temperature
(2-30°C). Store the bottle out of direct sunlight. Do
not use after expiry date. Do not touch any reagent area.
REGEANT AREA INFORMATION
Leukocytes
Normal urine specimens generally yield negative results;
positive results are clinically significant. Individually
observed “Trace” results may be of questionable
clinical significance; however ”Trace” results
observed repeatedly may be clinically significant. “Positive” results
may occasionally be found with random specimens from females
due to contamination of the specimen by vaginal discharge.
Higher glucose concentration (166mmol/L) or high specific
gravity may cause decreased test results. The presence of
cephalexin, cephalothin or high concentrations of oxalic
acid may also cause decreased test results. Tetracycline
may cause decreased reactivity and high levels of that drug
may cause a false negative reaction. Nitrofuration gives
a brown color to the urine that may mask the color reaction
on the reagent pad. Any substance that causes abnormal urine
color may obscure the color reaction.
Nitrite
This test is specific for nitrite in urine. Pink spots or
pink edges should not be interpreted as a positive re
sult. Any degree of uniform pink color development should
be interpreted as a positive nitrite test, suggesting the
presence of 1 ~ or more organisms per ml. but color development
Is not proportional to the number of bacteria present. A
negative result does not prove that there is no significant
bacteriuria. Prolonged urinary retention in bladder (4-8
hours) is essential in order to obtain an accurate result;
or lest is reduced for urine with high specific gravity.
Ascorbic acid concentrations of 1.42 mmol/L or greater may
cause false negative results with specimens containing small
amounts of nitrite ion (13µmol/L or less).
Urobilinogen
A result of 3.3µmol/L represents the transition from
normal to abnormal, and the patient and/or urine specimen
should be evaluated further. The reagent area may react with
substances known to interfere with Ehrlich’s reagent,
such as p-aminosalicylic acid and sulphonamides. Atypical
color reactions may be obtained in the presence of high concentrations
of paminobenzoic acid. False negative results may be obtained
if formalin is present. Highly colored substances, such as
azo dyes and riboflavin may mask color development on the
test area. Strip reactivity increases with temperature; the
optimum temperature is 22-26°C. The absence of urobilinogen
cannot be determined with the test.
Protein
The reaction is extremely sensitive to albumin. A ”Negative” result
does not rule out the presence of other proteins. Normally
no protein is detectable in urine by conventional methods
although a minute amount is excreted by the normal kidney.
A color matching any block greater than “Trace” indicates
significant proteinuria. For urine of high specific gravity,
highly buffered or alkaline urine, the lest area may most
closely match the “Trace” color block even though
only normal concentrations of protein are present. Further
evaluation is needed or “Trace” results. False
positive results may also be obtained by contamination of
the urine specimen with quatemary ammonium compounds or chilorhexidin
based disinfectants.
pH
The pH area measures pH value range of 5-8.5 visually and
5-8.5 instrumentally.
Blood
The significance of the “Trace” reaction may
vary among patients, and clinical judgment is required for
assessment in each individual case. Development of green
spots or green color on the reagent area within 60 seconds
indicates the need for further investigation. Blood is often
found in the urine of menstruating females. This test is
highly sensitive to hemoglobin and thus complements the microscopic
examination. The test is equally sensitive to myoglobin as
!o hemoglobin. Higher specific gravity or captopril may reduce
the reactivity of the blood test. Certain oxidizing contaminants,
such as hypochloritle or microbial perioxidase associated
with urinary tract infection may produce false positive results.
Specific Gravity
This test reflects the ion concentration of urine and correlates
well with the refractometric method. If pH is equal to or
greater than 6.5, then add 0.005 to SG obtained. Instrumental
readings are automatically adjusted for pH by the instrument.
The test is affected neither by certain nonionic urine constituents
such as glucose nor by the presence of radiopaque dye. Highly
buffered alkaline urines may cause low readings relative
to other methods. Higher specific gravity readings may be
obtained in the presence of moderate quantities (1 -7.5g/L)
of protein.
Ketones
The test reads with acetoacetic acid in urine. It does not
react with acetone or ß-hydroxybutyric acid. Some high
specific gravity/low pH urine may give reactions up to and
including “Trace”. Clinical judgment Is needed
to determine “Trace” results. Normal urine specimens
usually yield negative results. Detectable levels of ketone
may occur in urine during physiological stress conditions
such as fasting, pregnancy and frequent strenuous exercise.
In ketoacidosis, starvation or with other abnormalities of
carbohydrate or lipid metabolism, ketones may appear in the
urine in large amounts before serum ketone is elevated. False
positive results (Trace) may occur with highly pigmented
urine specimens or those containing large amounts of levodopa
metabolites, Compounds such as Mesna that contain sulfhydryl
groups may cause false positive results or an atypical color
reaction.
Bilirubin
Normally no bilirubin Is detected In the urine by even me
most sensitive methods. Even trace amounts of bilirubin are
sufficiently abnormal to require further investigation. Atypical
colors may indicate that bilirubin-derived bile pigments
are present in the urine sample end may be masking the bilirubin
reaction. These colors may indicate the urine specimen should
be tested further. Indican (indoxyl sulfate) can, produce
a yellow-orange to red color response which may interfere
with the bilirubin reading. Ascorbic acid concentrations
of 1.4 mmol/L or greater may cause false negatives.
Glucose
The test is specific for glucose. In dilute urines containing
less than 0.28mmolIL ascorbic arid, as little as 2-2mmol/L.
of glucose, may produce a color change that might be interpreted
as positive. Ascorbic concentrations of 2.84mmol/L or greater
and/or high ketone concentrations (4mmol/L) may give false
negatives for specimens containing small amount of glucose
(4-7mmol/L). The reactivity may also vary with temperature.
Small amount of glucose is normally excreted by the kidney.
These amounts are usually below me sensitivity of this test,
bill no occasion may produce a color between the “Negative” and
the 5.5mmol/L color blocks and that is interpreted by the
instrument as positive.
Ascorbic Acid
The test is based on the principle of Tillman’s reagent.
Ascorbic avid can reduce indicator and cause color changing,
from blue into orange. This test can be used to determine
ascorbic cod concentration in sample and decide ascorbic
acids interference.
SPECIFIC CHARACTERISTICS
In clinical specimens, the sensitivity depends upon several
factors; the variability of color perception, specific gravity,
pH, and the lighting conditions when the product is read
visually. Each color block or instrumental display value
represents range of values. Because of specimen and reading
variability, specimens with analytic concentrations that
tall between two levels may give results at either level.
Exact, agreement between visual results and instrumental
results may not be found because of the inherent differences
between the perception of the human eye and the optical system
of the instruments.
The following table lists the generally delectable levels
of analyses in contrived urine; however, concentrations may
be detected under certain conditions:
|
| |
| Reagent
Area |
Sensitivity |
Instrumental
Range |
Visual
Range |
| Glucose
(Glucose) |
4-7mol/L |
0-56mol/L |
0-111mol/L |
| Bilirubin
(Bilirubin) |
7-14µmol/L |
0-50µmol/L |
0-50µmol/L |
| Ketone
(Acetoacetic acid) |
0.5-1.0mmol/L |
0-8mmol/L |
0-16mmol/L |
| Blood
(Hemoglobin) |
0.015-0.04mg/dl |
0-2.0mg/dl |
0-2.0mg/dl |
| Protein
(Albumin) |
0.15-2.0g/L |
0-3.0g/L |
0-20.0g/L |
| Urobilinogen
(Urobilinogen) |
3-17µmol/L |
3-135µmol/L |
3-135µmol/L |
| Nitrite
(Nitrite ion) |
13-22µmol/L |
-~+ |
-~+ |
| Leukocytes
(White Cell) |
0.2-0.5mg/dl |
0-12mg/dl |
0-12mg/dl |
| pH |
-- |
5.0-8.5 |
5.0-8.5 |
| Specific
Gravity |
-- |
1.005-1.030 |
1.000-1.030 |
| Ascorbic
Acid |
0.5-0.6mmol/L |
-- |
0.5-5.0mmol/L |
|
| |
| INGREDIENTS
(100 Strips) |
| Glucose |
Glucose
oxidase |
3.5mg |
| |
Peroxidase |
0.6mg |
| |
Potassium
iodide |
6.5mg |
| Bilirubin |
2,4-DIchloioaniline
diazzonium salt |
2.2mg |
| Ketone |
Sodium
nitroprusside |
25mg |
| Specific
Gravity |
Bromothymol
blue |
0.3mg |
| |
Poly
(methyl vinyl ether maleic acid sodium
sail) |
15mg |
| pH |
Methyl
red |
0.05mg |
| |
Bromothymol
blue |
1.0mg |
| Blood |
Isopropylbenzidine
hydroperoxide |
18mg |
| |
3,3'-Dimethylbenzidine |
5.5mg |
| Protein |
Tetrabromphenol
blue |
0.30mg |
| Urobilinogen |
p-Dimethylaminobenzaldehyde |
1.5mg |
| Nitrite |
p-Arsanilic
acid |
6.8mg |
| |
N-(1-Naphthyl)
Ethylenediamine |
2.4mg |
| Leukocytes |
3-Indoly-phenol
ester |
6.0mg |
| |
Benzendiazonium
salt |
0.4mg |
| Shelf
Life |
24
MONTHS |
|
|
| |
|
|
|
| |
| |
| << Back |
| |
|