|
|
| |
|
| ::
REAGENT STRIPS FOR URINE ANALYSIS :: |
|
|
|
|
|
 |
| Item
Code: US-03 |
|
|
| HICKS
URISCREEN Reagent Strips (11 Parameters) |
| |
INTENDED
USE
Urianalysis Reagent Strips test for Ascorbic acid, Glucose, Bilirubin,
Ketone, Specific Gravity, Blood, pH, Protein, Nitrite, Leukocytes. The
strip may be lead visually or instrumentally, using the appropriate Urine
Chemistry Analyzers such as the Clintek family of readers. |
| |
| SPECIMEN
COLLECTION & PROCEDURE |
| 1. |
Use
a fresh urine specimen, less than 4 hours
old, and place it into a clean, dry container.
Do not centrifuge. Test within one hour.
If not possible, refrigerate and restore
to room temperature before testing.
|
| |
|
| 2. |
Remove
one strip from the bottle and replace the
cap lightly. Briefly (no longer than one
second) immerse alt reagent areas into
the specimen. Wipe off excess urine on
the rim of the container.
|
| |
|
| 3. |
Hold
strip in vertical position. Refer to the
bottle label for specific reagent areas
on the test strip. Compare the test areas
with the color scale on the label. Proper
reading times are critical for optimal
results. See each reagent time as indicated
on bottle label. Coloration appearing only
along the edges of the test, or developing
after more than two minutes, has no diagnostic
value.
|
|
| |
LIMITATION
As with all laboratory tests, definitive diagnostic or therapeutic decisions
should not be based on any single test result or method.
STORAGE
Do not remove desiccants from the bottle. Store at room temperature (2-30°C).
Store the bottle out of direct sunlight. Do not use after expiry date.
Do not touch any reagent area.
REGEANT AREA INFORMATION
Leukocytes
Normal urine specimens generally yield negative results; positive results
are clinically significant. Individually observed “Trace” results
may be of questionable clinical significance; however ”Trace” results
observed repeatedly may be clinically significant. “Positive” results
may occasionally be found with random specimens from females due to contamination
of the specimen by vaginal discharge. Higher glucose concentration (166mmol/L)
or high specific gravity may cause decreased test results. The presence
of cephalexin, cephalothin or high concentrations of oxalic acid may
also cause decreased test results. Tetracycline may cause decreased reactivity
and high levels of that drug may cause a false negative reaction. Nitrofuration
gives a brown color to the urine that may mask the color reaction on
the reagent pad. Any substance that causes abnormal urine color may obscure
the color reaction.
Nitrite
This test is specific for nitrite in urine. Pink spots or pink edges
should not be interpreted as a positive re
sult. Any degree of uniform pink color development should be interpreted
as a positive nitrite test, suggesting the presence of 1 ~ or more organisms
per ml. but color development Is not proportional to the number of bacteria
present. A negative result does not prove that there is no significant
bacteriuria. Prolonged urinary retention in bladder (4-8 hours) is essential
in order to obtain an accurate result; or lest is reduced for urine with
high specific gravity. Ascorbic acid concentrations of 1.42 mmol/L or
greater may cause false negative results with specimens containing small
amounts of nitrite ion (13µmol/L or less).
Urobilinogen
A result of 3.3µmol/L represents the transition from normal to
abnormal, and the patient and/or urine specimen should be evaluated further.
The reagent area may react with substances known to interfere with Ehrlich’s
reagent, such as p-aminosalicylic acid and sulphonamides. Atypical color
reactions may be obtained in the presence of high concentrations of paminobenzoic
acid. False negative results may be obtained if formalin is present.
Highly colored substances, such as azo dyes and riboflavin may mask color
development on the test area. Strip reactivity increases with temperature;
the optimum temperature is 22-26°C. The absence of urobilinogen cannot
be determined with the test.
Protein
The reaction is extremely sensitive to albumin. A ”Negative” result
does not rule out the presence of other proteins. Normally no protein
is detectable in urine by conventional methods although a minute amount
is excreted by the normal kidney. A color matching any block greater
than “Trace” indicates significant proteinuria. For urine
of high specific gravity, highly buffered or alkaline urine, the lest
area may most closely match the “Trace” color block even
though only normal concentrations of protein are present. Further evaluation
is needed or “Trace” results. False positive results may
also be obtained by contamination of the urine specimen with quatemary
ammonium compounds or chilorhexidin based disinfectants.
pH
The pH area measures pH value range of 5-8.5 visually and 5-8.5 instrumentally.
Blood
The significance of the “Trace” reaction may vary among patients,
and clinical judgment is required for assessment in each individual case.
Development of green spots or green color on the reagent area within
60 seconds indicates the need for further investigation. Blood is often
found in the urine of menstruating females. This test is highly sensitive
to hemoglobin and thus complements the microscopic examination. The test
is equally sensitive to myoglobin as !o hemoglobin. Higher specific gravity
or captopril may reduce the reactivity of the blood test. Certain oxidizing
contaminants, such as hypochloritle or microbial perioxidase associated
with urinary tract infection may produce false positive results.
Specific Gravity
This test reflects the ion concentration of urine and correlates well
with the refractometric method. If pH is equal to or greater than 6.5,
then add 0.005 to SG obtained. Instrumental readings are automatically
adjusted for pH by the instrument. The test is affected neither by certain
nonionic urine constituents such as glucose nor by the presence of radiopaque
dye. Highly buffered alkaline urines may cause low readings relative
to other methods. Higher specific gravity readings may be obtained in
the presence of moderate quantities (1 -7.5g/L) of protein.
Ketones
The test reads with acetoacetic acid in urine. It does not react with
acetone or ß-hydroxybutyric acid. Some high specific gravity/low
pH urine may give reactions up to and including “Trace”.
Clinical judgment Is needed to determine “Trace” results.
Normal urine specimens usually yield negative results. Detectable levels
of ketone may occur in urine during physiological stress conditions such
as fasting, pregnancy and frequent strenuous exercise. In ketoacidosis,
starvation or with other abnormalities of carbohydrate or lipid metabolism,
ketones may appear in the urine in large amounts before serum ketone
is elevated. False positive results (Trace) may occur with highly pigmented
urine specimens or those containing large amounts of levodopa metabolites,
Compounds such as Mesna that contain sulfhydryl groups may cause false
positive results or an atypical color reaction.
Bilirubin
Normally no bilirubin Is detected In the urine by even me most sensitive
methods. Even trace amounts of bilirubin are sufficiently abnormal to
require further investigation. Atypical colors may indicate that bilirubin-derived
bile pigments are present in the urine sample end may be masking the
bilirubin reaction. These colors may indicate the urine specimen should
be tested further. Indican (indoxyl sulfate) can, produce a yellow-orange
to red color response which may interfere with the bilirubin reading.
Ascorbic acid concentrations of 1.4 mmol/L or greater may cause false
negatives.
Glucose
The test is specific for glucose. In dilute urines containing less than
0.28mmolIL ascorbic arid, as little as 2-2mmol/L. of glucose, may produce
a color change that might be interpreted as positive. Ascorbic concentrations
of 2.84mmol/L or greater and/or high ketone concentrations (4mmol/L)
may give false negatives for specimens containing small amount of glucose
(4-7mmol/L). The reactivity may also vary with temperature. Small amount
of glucose is normally excreted by the kidney. These amounts are usually
below me sensitivity of this test, bill no occasion may produce a color
between the “Negative” and the 5.5mmol/L color blocks and
that is interpreted by the instrument as positive.
Ascorbic Acid
The test is based on the principle of Tillman’s reagent. Ascorbic
avid can reduce indicator and cause color changing, from blue into orange.
This test can be used to determine ascorbic cod concentration in sample
and decide ascorbic acids interference.
SPECIFIC CHARACTERISTICS
In clinical specimens, the sensitivity depends upon several factors;
the variability of color perception, specific gravity, pH, and the lighting
conditions when the product is read visually. Each color block or instrumental
display value represents range of values. Because of specimen and reading
variability, specimens with analytic concentrations that tall between
two levels may give results at either level. Exact, agreement between
visual results and instrumental results may not be found because of the
inherent differences between the perception of the human eye and the
optical system of the instruments.
The following table lists the generally delectable levels of analyses
in contrived urine; however, concentrations may be detected under certain
conditions:
|
| |
| Reagent
Area |
Sensitivity |
Instrumental
Range |
Visual
Range |
| Glucose
(Glucose) |
4-7mol/L |
0-56mol/L |
0-111mol/L |
| Bilirubin
(Bilirubin) |
7-14µmol/L |
0-50µmol/L |
0-50µmol/L |
| Ketone
(Acetoacetic acid) |
0.5-1.0mmol/L |
0-8mmol/L |
0-16mmol/L |
| Blood
(Hemoglobin) |
0.015-0.04mg/dl |
0-2.0mg/dl |
0-2.0mg/dl |
| Protein
(Albumin) |
0.15-2.0g/L |
0-3.0g/L |
0-20.0g/L |
| Urobilinogen
(Urobilinogen) |
3-17µmol/L |
3-135µmol/L |
3-135µmol/L |
| Nitrite
(Nitrite ion) |
13-22µmol/L |
-~+ |
-~+ |
| Leukocytes
(White Cell) |
0.2-0.5mg/dl |
0-12mg/dl |
0-12mg/dl |
| pH |
-- |
5.0-8.5 |
5.0-8.5 |
| Specific
Gravity |
-- |
1.005-1.030 |
1.000-1.030 |
| Ascorbic
Acid |
0.5-0.6mmol/L |
-- |
0.5-5.0mmol/L |
|
| |
| INGREDIENTS
(100 Strips) |
| Glucose |
Glucose
oxidase |
3.5mg |
| |
Peroxidase |
0.6mg |
| |
Potassium
iodide |
6.5mg |
| Bilirubin |
2,4-DIchloioaniline
diazzonium salt |
2.2mg |
| Ketone |
Sodium
nitroprusside |
25mg |
| Specific
Gravity |
Bromothymol
blue |
0.3mg |
| |
Poly
(methyl vinyl ether maleic acid sodium sail) |
15mg |
| pH |
Methyl
red |
0.05mg |
| |
Bromothymol
blue |
1.0mg |
| Blood |
Isopropylbenzidine
hydroperoxide |
18mg |
| |
3,3'-Dimethylbenzidine |
5.5mg |
| Protein |
Tetrabromphenol
blue |
0.30mg |
| Urobilinogen |
p-Dimethylaminobenzaldehyde |
1.5mg |
| Nitrite |
p-Arsanilic
acid |
6.8mg |
| |
N-(1-Naphthyl)
Ethylenediamine |
2.4mg |
| Leukocytes |
3-Indoly-phenol
ester |
6.0mg |
| |
Benzendiazonium
salt |
0.4mg |
| Shelf
Life |
24
MONTHS |
|
|
| |
|
|
|
| |
| |
| << Back |
| |
|